Chapter 1: CFX Automation System II Installation
| 7CFX Automation System II | 7
1. Place bubble level in the center of the CFX System(s) block (Figure 6). Note where the
bubble is within the level.
2. Lift the plate handler arm up high enough to access the tower opening. Place the level
inside the center of the plate handler tower (Figure 7). Compare the position of the
bubble to its position when placed on the CFX System block. Because the CFX Systems
do not have leveling feet, the most important thing is to level the plate handler relative
to the CFX System(s).
■n
If the bubble position differs significantly in the plate handler tower and CFX System(s),
proceed to step 3
■n
If the bubble position does not differ significantly when placed in the plate handler tower
and CFX System(s), proceed to step 6
Fig. 6. Bubble level in CFX System block.
Fig. 7. Bubble level in tower.