Manual
2.4.2 Maximizer Mixer
The Maximizer mixer provides the larger capacity required when the Maximizer is used. The Maximizer
mixer may be used with or without a mixer barrel: the mixer body, mixer top, and mixer barrel are provided
with the system; an optional mixer barrel extender is available. The mixer barrels fit between the mixer body
and the mixer top and are used for higher flow rates. Refer to the following table to determine the
appropriate mixer volume.
Table 2-11.
Mixer Barrels and Mixer Capacity for the Maximizer Mixer
Flow Rate Barrel Extension Capacity Assembly Screws
0.5 to 10 ml/min none 750 µl
1/4-20 x 1/2” (1.3 cm)
10 to 40 ml/min 5 ml Barrel Extender 5 ml 1/4-20 x 1 1/2” (3.8 cm)
40 to 80 ml/min 12 ml Barrel Extender 12 ml 1/4-20 x 3 1/4” (8.2 cm)
The procedure for changing the mixer capacity is provided in the following section.
Figure 2-3. Maximizer Mixer and Mixer Barrel Extender
MIXER
TOP
MIXER
BODY
12 ml MIXER
BARREL EXTENDER
OUTLET PORT
INLET PORT
5 ml MIXER
BARREL
EXTENDER
MAXIMIZER MIXER
DESCRIPTION OF BIOLOGIC DUOFLOW SYSTEMSYSTEM OVERVIEW
2-19