2 Bio-Scale MT Column Instruction Manual
1.3 Column Specifications
All wetted parts are bio-compatible (Tefzel, polypropylene,
polyphenyl sulfide, borosilicate glass and buna-N rubber) and
resistant to cleaning and sanitation solvents including 1.0 M NaOH
and 25% acetic acid. The column can be used with most common
aqueous buffers and certain organic solvents including alcohols
and acetonitrile. Do not use halogenated hydrocarbons, aromatic
solvents, or tetrahydrofuran (THF). A detailed solvent compatibility
table is given in Section 4.
Table 1. Column Characteristics.
MT2 MT5 MT10 MT20
Column
Volume, ml
1.9 – 2.3 4.6 – 5.7 9.5 – 11.3 19.4 – 21.9
Column
Dimension, mm
7 x 50 – 60 10 x 59 – 72 12 x 84 – 100 15 x 110 – 124
Maximum
Operating
Pressure, psi
1, 000 750 600 500
1145 psi = 1 MPa. 14.5 psi = 1 bar. Packing material may not withstand high
pressures. See packing material manufacturer’s specifications.