130
Users are required to enter their user name and password when they:
Start the application
Start a reading
Unlock the application
Sign a document, such as a Protocol or Results file
Different users may authenticate operations. For example, a Clinician
1-level user may be logged into Bio-Plex Manager, but a Supervisor-level user
may enter his user name and password to authenticate a particular reading.
Note that in this case, the Clinician 1-level user will remain logged onto the
application.
Your organization must establish its own guidelines for user authentication of
specific operations.
Electronic Records
Bio-Plex Manager
Security Edition supports the creation and control of secure
“electronic records,” as defined by 21 CFR Part 11. In Bio-Plex Manager, the
following are electronic records:
Protocol files
Results files
Calibration Log (included in the bioplexdata.mdb file)
Validation Log (included in the bioplexdata.mdb file)
Instrument Operations Log (included in the bioplexdata.mdb file)
These records are accessible only through Bio-Plex Manager and its
associated utilities.
File Security and Validation
The files used by Bio-Plex Manager Security Edition cannot be opened or
edited using other programs. Bio-Plex Manager checks the integrity and
validity of a file each time it opens. If a file becomes damaged or shows signs
of tampering with an outside application, it will not open in Bio-Plex Manager.
BioPlex_6.book Page 130 Friday, September 23, 2011 11:55 AM