Software Upgrade User guide
2 | An Overview of Bio-Plex Manager MP
12 | Bio-Plex Manager MP Software
After the routine has been successfully run, None is displayed indicating that
maintenance of the instrument is up to date.
Note: There may be times when you have to run more than one recommended
r
outine before maintenance is up to date. For example, Enhanced Maintenance
may be displayed as the recommended routine if calibration or verification fails.
After the Enhanced Maintenance routine is successfully completed, Daily Start
Up is displayed in the Recommended Routine section. Run the recommended
routines until None is displayed.
Tip: If calibration, verification, or both routines fail after successive attempts,
contact Bio-Rad T
echnical Support for assistance.
None is displayed when maintenance is up to date
The maintenance of the MAGPIX instrument is up to date and you can now create
and run your protocols.