User Manual

4. Always use forceps or wear gloves when handling membranes. Prewet the nitrocellulose
or Zeta-Probe
®
membrane by slowly sliding it at a 45° angle into wetting solution. Note:
PVDF membrane is not recommended. Wet nitrocellulose in 6x sodium, sodium citrate
(SSC) for nucleic acid applications, and in tris-buffered saline (TBS) for protein blotting.
Wet Zeta-Probe membrane in distilled water. See Sections 9 and 10 for solution
preparation. A 10 minute soak is recommended for complete wetting of the membrane
to insure proper drainage of solutions. Remove the membrane from the wetting solution.
Let the excess liquid drain from the membrane. (Touching the membrane to a sheet of
filter paper is a simple method for removing excess buffer.) Lay the membrane on the
gasket in the apparatus so that it covers all of the holes. The membrane should not
extend beyond the edge of the gasket after the Bio-Dot apparatus is assembled.
Remove any air bubbles trapped between the membrane and the gasket.
Note: PVDF membrane is not recommended.
5. Place the sample template on top of the membrane. The guide pins ensure that the
template will be properly aligned. Finger–tighten the four screws. When tightening the
screws, use a diagonal crossing pattern to ensure uniform application of pressure on
the membrane surface (see Figure 2).
Fig. 2. Diagonal crossing pattern for tightening screws in the Bio-Dot apparatus.
6. Attach a vacuum source (house vacuum or a vacuum pump) to the flow valve with a
waste trap set up and positioned between the vacuum outlet and the flow valve. Turn
on the vacuum and set the 3-way valve to apply vacuum to the apparatus (flow valve
setting 1, Figure 3).
7. With vacuum applied, repeat the tightening process using the diagonal crossing pattern.
Tightening while vacuum is applied ensures a tight seal, preventing cross contamination
between slots. Failure to tighten screws during application of vacuum prior to
starting the assay may lead to leaking between the wells.
8. Adjust the flow valve so that the vacuum manifold is open to air pressure (flow valve
setting 2, Figure 3). Apply 100 µl buffer to all 96 sample wells. Use of the 8-channel
pipet and buffer reservoirs (see Section 13 for ordering information) will simplify the
process of adding solutions to the Bio-Dot apparatus. Addition of buffer is necessary
to rehydrate the membrane following the vacuum procedure in step 7. If this step is
not performed prior to applying samples, assay results will show halos or weak
detection signal.
9. Gently remove the buffer from the wells by vacuum (flow valve setting 3, Figure 3).
Watch the sample wells. As soon as the buffer solution drains from all the wells,
adjust the flow valve so that the unit is exposed to air and disconnect the vacuum. At
this point, the unit is ready for sample application.
4