User Manual
Section 10
Troubleshooting Guide
I. Filtration Apparatus
1. Leakage or Cross-Well Contamination
a. Improper assembly. The screws must be retightened under vacuum following the
initial assembly.
b. Membrane is not properly rehydrated after assembly. Always rehydrate the
membrane prior to applying samples. Apply vacuum only until solutions are
removed from the sample wells, then disconnect the vacuum source.
2. No Filtration or Uneven Filtration Occurring
a. Macromolecular polymers, cellular debris, or dirt is plugging the membrane.
Centrifuge samples prior to application to remove particulates. Filter solution prior
to use to ensure removal of particulate material. Cover wells with Parafilm during
lengthy incubations.
b. Bubbles are obstructing the filtration. Use a needle to break any bubbles, being
careful not to puncture the membrane. Pipet liquid in the wells up and down to
displace bubbles.
c. The flow valve is positioned higher than the apparatus. The flow valve must be
lower than the level of the sample wells on the apparatus for proper drainage to
occur.
d. Improper blocking reagent is used. BSA is the recommended blocker for
nitrocellulose; gelatin will plug the apparatus, and no filtration will occur. The
Zeta-Probe
®
membrane, which requires more stringent blocking with BLOTTO or
with gelatin and MPO, should be removed from the Bio-Dot apparatus following
antigen immobilization and the rest of the assay should be conducted in a
separate container.
3. Halos
a. Membrane is not properly rehydrated before applying samples. Always rehydrate
membrane prior to applying any sample.
b. Excessive concentrations of sample are loaded. When too much sample is
present, wicking into the membrane around the well will occur. Use serial dilutions
of the samples to determine optimal amounts to load.
II. Poor Binding to Membrane
1. Nitrocellulose
a. DNA/RNA will only bind efficiently in 20x SSC or 1 M ammonium acetate. Use the
Zeta-Probe membrane as an alternative.
b. DNA must be single stranded and RNA must be denatured. DNA 500 bp may
not bind to nitrocellulose. Use the Zeta-Probe membrane as an alternative.
c. Mixed-ester cellulose binds DNA, RNA, and protein very poorly. Use Bio-Rad’s
pure nitrocellulose.
21