375 Marcus Boulevard • Hauppauge, NY 11788 • USA 631.273.0500 • Fax: 631.273.0771 e-mail: info@ultraviolet.com Extensive Product Information Available at: www.ultraviolet.com Document No.
TABLE OF CONTENTS SAFETY WARNINGS.....................................................................................................................................................3 SAFETY INSTRUCTIONS ............................................................................................................................................3 PRODUCT APPLICATION ...........................................................................................................................................
SAFETY WARNINGS • • All personnel should be alerted to the potential hazards indicated by the product safety labeling on this unit. The following conventions are used to indicate and classify precautions in this manual and on product safety labeling. Failure to observe precautions could result in injury to people or damage to property. ! This is the safety alert symbol. It is used to alert you to potential personal injury hazards.
PRODUCT APPLICATION CONSTRUCTION • • • • The water purifier is designed to mount horizontally. The water purifier's dual chamber heads are removable and may be rotated independently, which aids in the installation, maintenance or the retrofitting of an existing system. The water purifier's chamber, and chamber heads are electropolished and passivated type 316 stainless steel.
INSTALLATION UNPACKING • • • • • • Remove water purifier from shipping carton. Atlantic Ultraviolet Corporation takes all possible precautions when packaging this unit to prevent damage. Carefully inspect the water purifier, power cord and plug for damage, report all damage immediately. Do not assemble or operate if there is any damage, visible or suspected.
INSTALLATION 1. The water purifier should be mounted horizontally, on a flat dry surface. The water purifier should be secured using the mounting clips. The purifier should not be solely supported by its plumbing connections. 2. The water purifier must be connected to the cold water line only. POINT OF USE (POU) Ultraviolet Water Purifiers should be installed closest to the point of use.
LOCATION 1. 2. 3. 4. 5. 6. 7. The water purifier is intended for indoor use only. The water purifier should be protected from the elements and from temperatures below freezing. The ambient temperature, in the area surrounding the water purifier, should be between 33 F and 100 F. Electrical power supplied to the water purifier MUST match power requirements listed on the water purifier. Use of a voltage surge protector is recommended.
LAMP INSTALLATION OR REPLACEMENT Figure 3 - Lamp Replacement 1. 2. 3. 4. Step 2 - Remove End-Cap Step 3 - Remove Socket Step 4 - Remove Lamp Disconnect power to water purifier, and remove ground connector from the chamber. Remove easy-off end cap by pulling cap off static gland nut. Carefully withdraw lamp approximately 2 inches from chamber. While holding lamp end, remove lamp socket from exposed lamp end. Carefully withdraw lamp from chamber.
QUARTZ SLEEVE CLEANING OR REPLACEMENT Figure 4 - Quartz Sleeve Cleaning or Replacement Step 4 - Remove Gland Nuts Step 5 - Remove Washer & O-Ring Step 6 - Remove Quartz Sleeve 1. Disconnect power to water purifier, and remove ground connector from the chamber. 2. Shut off water supply to water purifier via inlet and outlet shut off valves and drain chamber, if possible. 3. Follow the steps in “LAMP INSTALLATION OR REPLACEMENT” to remove lamp. ! CAUTION: Lamp and quartz sleeve are easily damaged.
REPLACEMENT OF BROKEN QUARTZ SLEEVE Figure 5 - Chamber Head Removal 1. 2. Follow the steps in “Quartz Sleeve Cleaning or Replacement” to remove lamp and quartz sleeve. ! CAUTION: Considerable force maybe required to pull purifier from the mounting clips. Exercise extreme care when removing purifier. 12. Remove the water purifier from the mounting clips. Grasp the purifier chamber, and firmly pull each end free of the mounting clips.
TROUBLESHOOTING ! CAUTION: Always disconnect power to the water purifier, before performing any service. IMPORTANT: This unit is to be serviced ONLY by qualified, and appropriately licensed, personnel. Table 3 - Troubleshooting Problem Possible Cause Corrective Action Purifier not operating. No electrical power… Verify that the purifier is connected to a live power source. Cracked or broken Shut down purifier, drain, and replace quartz sleeve.
3-7/8" Height: 100 PSI 33° F - 100° F 14 Watts 100 PSI 33° F - 100° F Yes Power Consumption:d Max Operating Pressure: Ambient Temperature: Lamp Out Indicator: Yes 33° F - 100° F 100 PSI 25 Watts 0.21A 0.42A 120V 50/60Hz 3-7/8" 3-21/32" 24-7/8" Yes G24T6L 1 3/4" f NPT 4.0 BIO 4.0 Yes 33° F - 100° F 100 PSI 32 Watts 0.27A 0.54A 120V 50/60Hz 3-7/8" 3-21/32" 30-7/8" Yes G30T6L 1 3/4" f NPT 5.0 BIO 5.0 Yes 33° F - 100° F 100 PSI 39 Watts 0.325A 0.
OPTIONAL ACCESSORIES Optional accessories are available for operation at 120v 60Hz or 220v 50Hz, except Nylon solenoid valves. Please specify. Consult Factory for special input power requirements. Table 5 - Optional Accessories Optional Accessories Available for: Sentry™ Safety Sensor BIO 1.5 through BIO 12.0 Steralert TM Lamp Status Alarm BIO 1.5 through BIO 12.0 Time Delay Mechanism BIO 1.5 through BIO 12.
REPLACEMENT PARTS Bio-Logic ™ MODELS BIO-1.5 to BIO-12.
DESCRIPTION: Rubber O-ring, Static Gland Lamp Socket & Ballast, 120v 60Hz Lamp Socket & Ballast, 220-240v 50/60Hz Lamp Socket & Ballast, 12v DC Germicidal Lamp Static Gland Nut Teflon® Washer Rubber O-ring, Quartz Sleeve Quartz Sleeve Chamber Head, 2 ½" with Ground Terminal Chamber Clamp Chamber O-ring Chamber Tube Chamber Head, 2 ½" Easy-Off End Cap (no hole) Mounting Clip NO.
USER ASSISTANCE Atlantic Ultraviolet Corporation makes every effort to ensure that the Bio-Logic ™ Ultraviolet Water Purifiers are products of superior quality and workmanship. This manual describes the installation, operation and maintenance of the Bio-Logic ™ Ultraviolet Water Purifiers. Please read and become familiar with the contents of this manual before installing or using this unit.