Instruction manual

5.3.5.1 Collect All Mode
Collect All mode uses a single fraction size to collect fractions throughout a method. The units used for the
fraction size are either time (minutes) or volume (ml), depending upon the choice made when the method
was created. Table 18 describes how to program Collect All Mode.
Table 18.
Program Mode’s Fraction Collector Table: Collect All
Reference Description
Fraction Size: Enter the fraction size desired using the number
keys. The OK softkey confirms the fraction size entered. The
Cancel softkey reverts to the original or the default fraction
size. Both softkeys display the Fraction Collection Table for
Collect All mode.
Fraction Collection Table for Collect All mode: This table
shows the current fraction mode and the collection parameters.
To change the parameters press the Edit softkey. To display
the Method List press the OK softkey.
Note: To program the Pump Table or the Alarm Table without
returning to the Method List, press the Pump and Alarm keys
respectively while viewing the Fraction Collector Table.
64
System Operation
Method: Volume: 1.00 ml
Enter Fraction size: 1.00ml
CANCEL OK
F2
F3
F3
Method: Volume: 1.00 ml
COLLECT ALL MODE
Fraction size: 1.00ml
EDIT OK
F3
F2
Figure 26. Programming: the Fraction Collector Table’s Collect All
(Screen displays are keyed to the discussion in Table 18)
Method: name Volume: 1.00 ml
COLLECT ALL MODE
Fraction Size: 1.00 ml
EDIT OK
Method: name Volume: 1.00 ml
Select a Fraction Collector Mode
ALL THRESHOLD WINDOWS THRES+WIN
Method: name Volume: 1.00 ml
Enter Fraction size: 1.00ml
CANCEL OK
F1 F1a
F2
F2a
(return to
Method List)
(return
to F2)
To remove Method's fraction collection
mode; no fraction collection will occur
when the Method is run.
F2
LEGEND: SELECT COLLECTION MODE
COLLECT ALL MODE: ENTER FRACTION SIZE
FRACTION COLLECTOR TABLE FOR COLLECT ALL
Method: name Volume: 2.00 ml
FC Mode: Collect all
Estimated Number of Fractions: 20
EDIT ERASE NEW MODE OK
(return
to F1)
F2a
Displays
Method List