Specifications

14
Q
UARTZ SLEEVE CLEANING OR REPLACEMENT
Step 4 - Remove Gland Nuts
Step 6 - Remove Quartz Sleeve
Step 5 - Remove Washer & O-Ring
1. Disconnect power to water purifier.
For 120v 60Hz version using ballast no. 10-0010B, remove the ground connector from the chamber,
and then go to next step.
For 230v 50/60Hz version using ballast no. 10-0504A, go to next step.
2. Shut off water supply to water purifier via inlet and outlet shut off valves and drain chamber, if possible.
3. Follow the steps in “LAMP INSTALLATION OR REPLACEMENT” to remove lamp.
CAUTION: Lamp and quartz sleeve are easily damaged. Exercise care when handling.
4. Unscrew static gland nuts from each end of the chamber. Avoid striking quartz sleeve with static gland nut.
5. Remove Teflon
®
washer and O-ring from both ends of quartz sleeve. Teflon
®
washer will sometimes remain
within the static gland nut. If so, remove Teflon
®
washer from static gland nut before proceeding.
6. Carefully remove quartz sleeve from chamber. NOTE: It is advisable to support the quartz sleeve on the backside
with your finger so that it does not drop to the bottom of the chamber as it slides into the chamber.
7. Once the quartz sleeve is removed, clean with alcohol or a detergent. Stubborn stains usually can be removed with
a dilute hydrochloric acid. NOTE
: Follow all manufacturer's instructions and precautions when handling
chemicals.
8. Reassemble in reverse order. Make sure the quartz sleeve protrudes an equal distance past each threaded nipple.
Be sure O-rings are placed on quartz sleeve before Teflon
®
washer.
9. Tighten static gland nuts firmly by hand only, DO NOT USE HAND TOOLS. Tightening with hand tools is likely
to cause quartz sleeve to break.
10. When all connections are complete allow water to enter the water purifier at a low flow rate until the purifier is
pressurized. With the purifier pressurized, it should be checked for leaks.
11. If no leaks occur, reinstall lamp,
following the steps in “LAMP INSTALLATION OR REPLACEMENT” section.
WARNING: Germicidal ultraviolet rays are harmful to eyes and skin. Do not restore power to
water purifier until lamp and both easy-off end caps have been properly reinstalled.
!
!
Figure 10 - Quartz Sleeve Cleaning or Replacement
Bio-Logic
®
BIO-1.5 Pure Water Pack
TM
120v 50/60Hz (ballast no. 10-0010B)
Figure 11 - Quartz Sleeve Cleaning or Replacement
Bio-Logic
®
BIO-1.5 Pure Water Pack
TM
230v 50/60Hz (ballast no. 10-0504A)