Specifications

13
LAMP INSTALLATION OR REPLACEMENT: BIO-LOGIC
®
BIO-1.5 PURE WATER PACK
TM
230V 50/60HZ USING BALLAST NO.10-0504A
Figure 8 - Lamp Replacement for Bio-Logic
®
BIO-1.5 Pure Water Pack
TM
230v 50/60Hz (ballast no. 10-0504A)
1. Disconnect power to water purifier.
2. Remove easy-off end cap by pulling cap off static gland nut.
3. Pull lamp cord and carefully withdraw lamp approximately 2 inches from chamber. While holding lamp end,
remove lamp socket from exposed lamp end.
4. Carefully withdraw lamp from chamber. Be sure to withdraw lamp straight out without angling until completely
clear of quartz sleeve.
CAUTION: Lamp and quartz sleeve are easily damaged. Exercise care when handling.
5. Reinstall lamp in reverse order.
WARNING:
Germicidal ultraviolet rays are harmful to eyes and skin. Do not restore power to
water purifier until lamp and easy-off end caps have been properly reinstalled.
!
!
Step 2 - Remove End-Cap
Step 3 - Remove Socket
Step 4 - Remove Lamp
Step 1 - Disconnect Power
Disconnect Power