LAMP INSTALLATION OR REPLACEMENT: BIO-LOGIC
®
BIO-1.5 PURE WATER PACK
TM
120V 50/60HZ USING BALLAST NO. 10-0010B
1. Disconnect power to Pure Water Pack
TM
, and remove ground
connector from the Bio-Logic
®
chamber.
2. Remove easy-off end cap by pulling cap off static gland nut.
3. Carefully withdraw lamp approximately 2" from chamber.
4. While holding lamp end, remove lamp socket from exposed lamp end.
5. Carefully withdraw lamp from chamber. Be sure to withdraw lamp
straight out without angling until completely clear of quartz sleeve.
CAUTION: Lamp and quartz sleeve are easily damaged.
Exercise care when handling.
6. Reinstall lamp in reverse order.
WARNING: Germicidal ultraviolet rays are harmful to
eyes and skin. Do not restore power to water purifier until lamp
and both easy-off end caps have been properly reinstalled.
7. Reattach the ground connector to the grounding tab of the chamber,
and restore power to the system.
DISPOSAL OF MERCURY ADDED LAMPS
Germicidal ultraviolet lamps, like standard fluorescent lamps contain small
amounts of mercury. Mercury added lamps should not be placed in the
trash. Dispose of properly.
For further information regarding the disposal and recycling of lamps con-
taining mercury, along with Federal and State requirements visit
http://www.lamprecycle.org. Product Data Sheets for germicidal ultraviolet
lamps can be found in the “PDF Library for Ster-L-Ray
®
Germicidal
Ultraviolet Lamps” section of http://www.ultraviolet.com.
!
!
Step 1 - Disconnect Power and
Remove Ground Connector
Step 2 - Remove End-Cap
Step 3 - Remove Socket
Step 4 - Remove Lamp
Figure 7 - Lamp Replacement