User's Guide

This document is subject to change without prior notification.
Binsfeld Engineering Inc. | 231-334-4383 www.binsfeld.com 869610-9_A2 Pg. 13
2.2.5.3 Standby Mode
This is the TX20K low power, sleep mode. It is used to conserve battery
power between test sessions. No torque or status data is transmitted
while in Standby Mode. The TX20K wakes from this mode every 15
seconds to see if the RX20K is sending Wake commands. If not, the
TX20K goes back to sleep. If the TX20K receives a Wake command
from the RX20K, it will wake and operate at same Block Mode
baud/sample rate as when it was put to sleep. Resetting power of the
TX20K will take it out of Standby Mode and back into Block Mode
operation.
2.2.6 Signal Accuracy
Below are the TX20K’s maximum signal errors at 25
˚C along with the
temperature coefficients for each range setting.
Table 5, TX20K signal accuracy
Input
Range
mV/V
Zero Error
% Full Scale
Scale Error
% Reading
Zero TC
ppm/°C
Scale TC
ppm/°C
0.2
0.02
0.25
50
30
0.5
0.02
0.25
30
25
1
0.15
0.20
25
20
2
0.15
0.20
20
20
5
0.15
0.20
20
20
10
0.15
0.20
20
20
20
0.15
0.20
20
20