Technical data
C 150 (E2) 12/2012 page 15/90
The CO
2
incubators were produced in accordance with VDE regulations and were routinely tested in ac-
cordance to VDE 0411-1 (IEC 61010-1).
CAUTION
The glass door and the inner chamber will become hot during operation.
Danger of burning.
∅ Do NOT touch the glass door, the inner surfaces or the charging material during opera-
tion.
WARNING
Stability hazard.
Danger of injury.
Damage to the unit and the charging material.
Housing cover breakaway.
∅ Do NOT climb on the lower housing cover.
∅ Do NOT load the lower housing cover with heavy objects while the unit door is open.
1.6 Precautions when working with CO
2
gas
Carbon dioxide (CO
2
) in high concentrations is hazardous to health. It is colorless and almost odorless
and therefore practically imperceptible. Vent out any CO
2
gas that may escape via good room ventilation
or a suitable connection to an exhaust system. We recommend installing a CO
2
warning system.
WARNING
High concentration of CO
2
(> 4 Vol.-%).
Risk of death by suffocation.
Danger of poisoning.
∅ Do NOT set up units in non-ventilated recesses
Ensure technical ventilation measures
Observe the relevant regulations for handling CO
2
.