Technical data

BD / ED / FD 01/2006 page 36/49
11.2 Decommissioning
Switch off ED units sizes 400 and 720 at the main switch (10). Disconnect the unit from the mains.
When switching off the main switch ON / OFF (10), the stored parameters remain saved.
Temporal decommissioning: See indications for appropriate storage, chap. 4.3.
Final decommissioning: Dispose of the unit as described in chap. 11.3.
11.3 Disposal of the unit
CAUTION
Alteration of the environment.
For final decommissioning and disposal of the oven, please contact the BINDER
service.
Observe the regulations under public law for appropriate disposal protecting the
environment.
The main board of the oven includes a lithium cell. Please dispose of it according to the national
regulations.
12. Technical description
12.1 Factory calibration and adjustment
This unit was calibrated and adjusted in factory. Calibration and adjustment were performed using
standardized test instructions, according to the QM-system of DIN EN ISO 9001 applied by BINDER
(certified since December 1996 by TÜV CERT under registration number 70 100 M 926). All test
equipment used is subject to the administration of measurement and test equipment that is also
constituent part of the BINDER QM-systems of DIN EN ISO 9001. They are controlled and calibrated in
relation to a DKD-Standard on regular intervals.
Adjustment in factory in the center of the usable volume BD: 37°C.
Adjustment in factory in the center of the usable volume ED and FD: 150°C.
12.2 Definition of usable space
The usable volume illustrated below is calculated as follows:
b
c
c
a
b
a
C
B
A
A, B, C = Internal dimensions (W, H, D)
a, b, c = Wall clearances
a = 0.1 x A
b = 0.1x B
c = 0.1 x C
V
USE
= (A - 2a) x (B - 2b) x (C - 2c)
Figure 10: Determination of the useable volume
The technical data refer to the so defined usable space.