Technical data

BD / ED / FD 01/2006 page 33/49
9.5 Water protected internal socket (option BD)
The internal socket is splash proof.
IP system of protection 65 230 V 1N ~ 50-60 Hz
Charge max. 500 W
Maximum permitted operating temperature with this option: 50°C.
WARNING
Exceeding of the permitted maximum temperature.
Electrical hazard.
Danger of life.
Damage of the internal socket.
Set the mechanical thermostat class 3.1 to 50°C.
Do NOT exceed the temperature set-point of 50°C.
Heat emission of electrical devices connected inside the chamber may modify the
temperature range.
CAUTION
Risk of short circuit.
Damage of the unit.
Use delivered plug only (IP protection type 66). Plug-in the plug and turn it to secure.
If the socket is not used, close the lift-up lid and turn it for securing.
10. Maintenance, cleaning, and service
10.1 Maintenance intervals, service
DANGER
Electrical hazard.
Danger of life.
Put off-circuit the unit before conducting maintenance work. Pull the power plug.
The unit must NOT become wet during operation or maintenance works.
Have all maintenance work conducted by professional electricians or experts
authorized by BINDER.
Have conducted regular maintenance work at least once a year.
Change the door gasket in cold condition only. Otherwise the door gasket will be damaged.