Technical data

BD / ED / FD 01/2006 page 23/49
The display shows alternately “SP” and in the entry level the previous temperature set-point
(example: 60°C):
3. With the buttons enter a set-point value between 0 and 300.
The desired temperature set-point can be selected in a temperature range from 5°C above
room temperature up to 300°C.
4. Wait 2
seconds until the entered temperature value is taken over (display flashing once).
5. Press
button to return to normal display / actual value display (automatically after 60 seconds).
7.3 Time functions: Continuous operation and Timer operation
Press the time management button .
The timer indicates its current time function. There are two possible time functions:
Continuous operation
The display shows alternately “t1” (time function) and the time function “Continuous operation” “t inf”:
The heating and fan (with FD) are permanently active, independent of the timer setting.
Timer operation
The display shows alternately “t1” (time function) and the running-down time or “tOff”:
or
Remaining time (example: 28 Min.) – Timer running
down
Heating and fan (with FD) are active until the timer has
run-down.
Timer not programmed or run-down “t off”
If the timer has run-down, heating and fan (with FD) are
permanently off.
Press
button to return to normal display (actual value display) (automatically after 60 seconds).