Datasheet

BET-1000M
100 Watt
Desktop adapter for medical applications
Class 2 equipment – without PE
Compliant with ErP and CEC level V
Input voltage 90…264 V AC
Input frequency 47…63 Hz
Inrush current 100 A max (240 V AC)
Efficiency 88…89 %
Idle consumption ≤0.5 W
Hold up time 16 ms / 115 VAC
Protection Short circuit protection: Switch off with auto recovery
Overvoltage protection
Insulation voltage Input / Output 4675 V AC (2 x MOPP)
Line regulation ±1 % at rated load and input voltage change of 100…240 VAC
Load regulation ≤ ±4 %
Leakage current <100 µA touch current (264 V AC / 63 Hz)
Safety / EMC TÜV (IEC/EN60601-1 3rd Edition), UL (UL60601-1 3rd Edition), CB, CE
Temperature Operating: -20…+60 °C / Storage: -20…+85 °C
Derating In the range of +40…+60 °C, 2.5 % / °C
Max. operation altitude 3000 m
MTBF >165 000 h according to MIL-HDBK-217F at +25 °C
Humidity Operating: 10…85 % RH, non-condensing / Storage: 10…90 % RH, non-condensing
Dimensions (W x D x H) 58 x 142 x 37 mm ±0.5 mm
Weight (net) 0.485 kg
Product specific data
ECO standard ErP directive 2009/125/EG, CEC level V
The series BET-1000M desktop adapters supply power to medical
devices. They are internationally certified according to IEC /
EN /
UL 60601-1 (3rd Edition) standards.
The desktop adapters are de-
signed for many years of reliable operation
.
Robustly constructed,
the adapters are also equipped with top quality components. The
entire closed plastic case can easily be cleaned and disinfected.
0%
25%
50%
75%
100%
0 10 20 30 40 50 60
Ambient Temperature(˚C)
20LFM(Natural Convection)
-20
-10
Derating
Cling pad, Art. No. KP-01
Te c h n i c a l d a t a
Extended
Extended
Fanless
Fanless
Eco
Eco
Medical
Medical
Delivery
5
YEARS
Delivery
5
YEARS
Medical applications
Desktop adapter
Bicker Elektronik GmbH | 86609 Donauwörth | Germany
Tel. +49 906 70595-0 | info@bicker.de | www.bicker.de