Operator`s manual
26
Communication Menu Configuration
88
SUM.CHK
Press ‘RESET/ENTER’ 14)
Display shows "SUM.CHk" Check Sum setting submenu. If necessary
to select the other available following Bus Format Setting submenus:
"LN.FEEd", "ECHO", "StNd", "MODE", and "HNd.SHk", use or press the
‘MENU’ button.
Press ‘RESET/ENTER’ 15) Display flashes "NO" No (checksum excluded) option as default or
previous selection.
Press ‘䊱/MAX’ 16)
Scroll through or select either two available options: "NO" to exclude or
"yES" to include the checksum with Input Reading Data.
LN.FEEd
Press ‘RESET/ENTER’ 17) Display shows "StOREd" stored message momentarily and then
advances to the "LN.FEEd" Line Feed setting submenu.
• Line Feed setting determines if data sent from the controller will have a
Line Feed appended to the end or following all <CR>’s – useful for
viewing or logging results on separate lines when displayed on
communications software at a computer.
Press ‘RESET/ENTER’ 18) Display flashes "NO" No (Line Feeds excluded) option as default or
previous selection.
Press ‘䊱/MAX’ 19)
Scroll through or select either two available options: "NO" to exclude or
"yES" to include the line feeds with Reading Data.
ECHO
Press ‘RESET/ENTER’ 20)
Display shows "StOREd" stored message momentarily and then
advances to the "ECHO" Line Feed setting submenu.
• When valid commands are sent to the instrument, this determines
whether the command will be echoed to the Serial Bus. Use of "ECHO"
is recommended in most situations
, especially to help verify that data
was received and recognized by the controller.
Press ‘RESET/ENTER’ 21) Display flashes "yES" Yes (enable Echo responding from controller)
option as default or previous selection.
Press ‘䊱/MAX’ 22) Scroll through or select either two available options: "yES" to enable
Echo response or "NO" to disable Echo response from Controller to Serial
Bus.