Operator`s manual

12
Input Type Selection
44
12 INPUT TYPE SELECTION: (INPUt)
12.1 INPUT TYPE: (THERMOCOUPLE)
INPUt ENTER INPUT TYPE MENU:
Press ‘MENU 1) Display shows "INPUt" Input Menu.
Press ‘RESET/ENTER 2)
Display flashes "VOLt" or previous stored input type.
Press ‘/MAX 3) Scroll through the available following input type selection list. Display
flashes:
"VOLt" (DC Voltage)
"CURRNt" (DC Current)
"bRIdGE" (Bridge)
"POt" (Potentiometer)
"tC" (Thermocouple)
"Rtd" (RTD).
If the displayed input type is "tC", press ‘RESET/ENTER’ to skip to Step 6
("tC" stops flashing).
t.C "t.C" THERMOCOUPLE SUBMENU:
Press ‘/MAX 4) Scroll through the available selection to "tC" (flashing).
Press ‘RESET/ENTER 5) Display shows "tC" again, but not flashing.
Press ‘RESET/ENTER 6) Display flashes previous thermocouple type selection. ie. "J tC" as
default, (see Table below for types). If the display matches your choice of
Thermocouple. Go to Step 8.
Press ‘/MAX 7) Scroll through the available thermocouple types to the selection of your
choice.
Press ‘RESET/ENTER 8) Display shows "StOREd" stored message momentarily and then
advances to the "RdG.CNF" Reading Configuration Menu.
Thermocouple
Types: J
K T E N DIN J R S B
Display: J tC k tC t tC E tC N tC dNJ tC R tC S tC b tC