Troubleshooting
32
“I OVSC”
POSSIBLE CAUSE: The input scale and/or offset values chosen are large enough to
drive the display into numerical overflow.
TO CORRECT:
Reduce the input and/or the input scaling/offset. Refer to IN.SC.OF
in Section 16.4.
“R OVSC”
POSSIBLE CAUSE: The reading scale and/or offset values chosen are large enough to
drive the display into numerical overflow.
TO CORRECT: Reduce the reading scale/offset and/or move the active decimal
point to the right. Reducing input scale/offset is not required,
because the input overscale message has higher priority than this
message and would be displayed if there were an input overscale.
“CB OVF”
POSSIBLE CAUSE: CNT BY (count by value) has been changed causing the display to
round up to a numerical overflow.
TO CORRECT: Reduce the CNT BYcount by value to 001. If you have an active
decimal point selected, move the decimal point one or more
positions to the right.
“SERIAL”
POSSIBLE CAUSE: A configuring change has been attempted via the front panel
buttons while the serial communications port is actively
communicating with the meter.
TO CORRECT: Either disconnect the serial cable, or press ‘MENU’ for configuration
of the meter.