Communication Menu Configuration
94
SEPARA
Press ‘RESET/ENTER 51) Display shows "StOREd" stored message momentarily and then
advances to the "SEPARA" Separator Selection submenu that determines
whether different data types included in Data String and sent from the
device will be separated by spaces or by Carriage Returns.
Press ‘RESET/ENTER 52) Display flashes "SPACE" Space separator as default or previous
selection.
Press ‘/MAX 53) Scroll and select "SPACE" or "CR" <CR> Carriage Return Option for
your choice of Data String Format.
UNIt
Press ‘RESET/ENTER 54) Display shows "StOREd" stored message momentarily and then
advances to the "UNIt" Unit of Measure submenu that allows user to
include user-Defined unit of measurement in Data String Format (For detail
and description of how to create your own units of measurement, refer to
Serial Communication Operation Manual).
Press ‘RESET/ENTER 55) Display flashes "NO" No (excluded as default) option or previous
selection.
Press ‘/MAX 56) Scroll and select "NO" to exclude or "yES" to include your own defined
unit of measure in the Data String for your choice of Data Format.
26.4. MOdbUS Enter MODBUS PROTOCOL SUBMENU
Press ‘RESET/ENTER 57) Display shows "StOREd" stored message momentarily and then
advances to the "MOdbUS" Modbus Protocol submenu.
Press ‘RESET/ENTER 58) Display flashes "NO" No option as default as Modbus Protocol
disabled or previous selection.
Press ‘/MAX 59) Scroll and select "NO" to disable or "yES" to enable Modbus protocol.
If Modbus Protocol was selected, the following Communications
Parameters must be set as: No Parity, 8-bit Data Bit, 1-Stop Bit.
Do not attempt to change these parameters
.