Specifications

Quick-Start Operating Guide
Humex 2™ Inline Humidification System
© Copyright 2012 Terra Universal Inc. All rights reserved. • Revised April 2012 Document No. 1800-69v2.0tm
Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020
7
Prg Purge Delay: The Dual Purge System senses the internal cabinet pressure and activates a high-flow
purge when this pressure falls below a set level. The “Prg” setting lets you specify the number of
seconds the high purge remains active after internal glove box pressure level is restored (default: 60).
door Number of seconds a door must remain open to activate the OPEN DOOR alarm (default: 60). Inactive
in Glove Box applications.
bEEP Enables and disables beeper (default: ON).
Glo Activates GLOVE BOX control mode: press UP to turn ON and proceed to glove box pressure setting;
DOWN to turn OFF (default: ON).
Press In GLOVE BOX mode, this setting lets you adjust
the pressure (milli-inches WG) at which the high
purge is activated (default: 0.2).
Done Settings are complete; press SET once more to
begin operation.
Make a mistake? No Problem! To reset the system, turn the
system OFF while in setup mode and restart while holding the
SET button.
Humex 2™ Operation with the Dual Purge™ System
3.6 Operation Display
Functions: After completing initial programming, press the
specified button to view these operating conditions:
High Purge Bypass Press and hold SET
Humex 2™/Dual Purge™ System Operation
Solenoid
Heater
wires
Heater
Vaporization
Chamber
Humex 2 Interior
(viewed from left side)