Specifications

Quick-Start Operating Guide
Humex 2™ Inline Humidification System
© Copyright 2012 Terra Universal Inc. All rights reserved. • Revised April 2012 Document No. 1800-69v2.0tm
Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020
3
When the set point is attained, the Humex deactivates the high flow (as well as the Humex humidification system) to
conserve on nitrogen. In this low-flow purge, the flow rate is restricted by the Dual Purge flowmeter.
Refer to Operating Manual No. 1800-42 or 1800-13 for further information on Terra’s Glove Box Dual Purge System.
3.0 Set Up and Operation: Humex 2™ without Dual Purge
Humex 2™ Set Up
Includes: Humex 2™ controller (Part No. 9081-02), humidity sensor (Part No. 9500-02A), 120VAC power cord, 12DC
power supply and cord, pinch valve assembly, water filter, humidity sensor cable, Door Switch jumper (installed on
rear panel), 0-20 SCFH flowmeter, .25" clear urethane tubing. Refer to illustration on page 2.
2.1. Carefully unpack the unit and check for any visible damage or missing parts. Any damage should be reported
immediately to the shipping company.
2.2. Make sure that your glove box is placed on a level, stable surface, away from heat or chemicals that could
damage it.
2.3. Using 0.25" clear urethane tubing, connect your in-house gas supply to the “Gas In” bulkhead on the rear panel
of the Humex 2 by loosening the Poly-Tite collar, inserting the tubing, and tightening the collar.
2.4. Use another length of 1/4" tubing to connect “Gas Out” on the rear panel to the gas inlet of the enclosure to be
monitored.
2.5. Connect the pinch valve assembly into the exit side of the water filter and the “Water” port on the rear panel.
2.6. Connect the water filter inlet side to a water source.
To Water Filter/Pinch
Valve Assembly
1/4"-dia. tubing
To 12VDC Power
Supply (Phone
jack connector)
Humidity Sensor
Cable (3-pin DIN
connector)
120VAC Power
Regulated Gas
Supply (35 psi;
15 20 SCFH);
1/4"-dia. tubing
Set Up for Humex 2™