Operation Manual

29
4. Maintenance
4.1 Replacing fuses
- F1
M 0.315 A / 250 V, 20 x 5 mm, T.Nr.: 757211
ThisfuseprotectsinternalcircuitryoflowΩresistancefunctioniftestprobesareconnectedto
the mains supply voltage by mistake.
- F2, F3
F 4 A / 500 V, 32 x 6.3 mm, T.Nr.: 757212
General input protection fuses of test terminals L/L1 and N/L2.
Warning:
- Disconnect any measuring accessory and power off the BENNING IT 120 B before
opening battery/fuse compartment cover, hazardous voltage inside.
- Replace blown fuse with original type only, otherwise the BENNING IT 120 B may be
damaged and/or operator’s safety impaired.
Battery and fuse compartment
Legend:
1 Fuse F1
2 Fuse F2
3 Fuse F3
4 Serial number label, a further serial number label is located outside next to the information label
5 Batterycells(sizeAA)
6 Battery holder
Maintenance Replacing fuses