Operation Manual

09/ 2013
BENNING CM 7
12
6. Environment conditions:
- The BENNING CM 7 is intended for making measurements in dry
environment.
- Maximum barometric elevation for making measurements: 2000 m,
- Overvoltage category/ Siting category:
IEC 60664-1/ IEC 61010-1 → 600 V category IV, 1000 V category III
- Contamination class: 2,
- Protection Class: IP 30 (DIN VDE 0470-1 IEC/ EN 60529)
IP 30 means: Protection against access to dangerous parts and protection
against solid impurities of a diameter > 2.5 mm, (3 - first index). No protec-
tion against water, (0 - second index).
- Operating temperature and relative humidity:
For operating temperature from 0 °C to 30 °C: relative humidity less than 80 %
For operating temperatures from 31 °C to 40 °C: relative humidity less than 75 %
F
or operating temperature from 41 °C to 50 °C: relative humidity less than 45 %
- Storage temperature: The BENNING CM 7 can be stored at any tempera-
ture in the range from - 20 °C to + 60 °C (relative humidity from 0 to 80 %).
The battery should be taken out of the instrument for storage.
7. Electrical specications
Note: The measuring precision is specified as the sum of
- a relative fraction of the measured value and
- a number of digits (counting steps of the least significant digit).
This specified measuring precision is valid for temperatures in the range from
18 °C to 28 °C and relative humidity less than 80 %.
7.1 Direct voltage ranges
The input resistance amounts to 1 M
Meas. range Resolution Meas. precision Overload protection
400 V 0,1 V ± (0.7 % of the measuring value + 2 digits) 750 V
eff
1000 V 1 V ± (0.7 % of the measuring value + 2 digits) 750 V
eff
7.2 Alternating voltage ranges
The input resistance amounts to 1 M in parallel 100 pF.
Measuring
range
Resolution
Meas. precision *
1
*
2
within the
frequency range 50 Hz - 500 Hz
Overload
protection
400 V 0,1 V ± (1 % of the measuring value + 5 digits) 750 V
eff
750 V 1 V ± (1 % of the measuring value + 5 digits) 750 V
eff
*
1
The measuring value is gained and indicated as effective value (True RMS,
AC coupling). The measuring accuracy is specified for sinusoidal curves
and applies to the final value of the measuring range as well as for non-
sinusoidal curves up to 50 % of the final value of the measuring range.
In case of non-sinusoidal curves, the indicating value becomes inaccurate.
Thus, an additional error occurs for the following crest factors:
crest factor from 1.4 to 2.0 additional error + 1 %
crest factor from 2.0 to 2.5 additional error + 2.5 %
crest factor from 2.5 to 3.0 additional error + 4 %
*
2
additionally ± 4 digits for measuring values < 15 % of the final value of the
measuring range
7.3 Direct current (DC) ranges
Meas. range Resolution Measuring accuracy Overload protection
0 - 200 A 0,1 A ± (2,9 % of the measuring value + 3 A) 1000 A
eff
200 - 400 A 0,1 A ± (1,9 % of the measuring value + 2 A) 1000 A
eff
400 - 1000 A 1 A ± (2,9 % of the measuring value + 5 A) 1000 A
eff
The indicated accuracy is specified for conductors which are gripped by means
of the measuring clamp in the middle (see Fig. 4 Direct/ alternating current
measurement). For conductors which are not gripped in the middle, an addi-
tional error of 1 % of the indicating value has to be considered.
Error according to remanence: 1 % (repeating measurement)
7.4 Alternating current ranges
Measuring
range
Resolution
Meas. precision *
1*2
within the
frequency range 50 Hz - 400 Hz
Overload
protection
0 - 200 A 0,1 A ± (2.9 % of the measuring value + 3 A) 1000 A
eff