Instructions

Operation
5.3 Voltage or frequency measurement
5191 / 07/2021 en BENNING CM 2-1 29
Figure11: DC voltage measurement
Procedure
1. Set the rotary switch of the device to switch position “Hz ” or “Volt Sense/ Lo‑Z Auto‑V”.
“Hz ”: Press the “ACHFR/ DCAZERO” key to set the desired measuring function
(frequency, AC or DC voltage).
“Volt Sense/ Lo‑Z Auto‑V”: The corresponding measuring function (AC or DC voltage)
and the optimum measuring range are set automatically. The input resistance is
reduced to approx. 3kΩ in order to suppress inductive and capacitive voltages (reactive
voltages).
2. Connect the safety measuring lines to the device [}page27].
3. Bring the safety measuring lines into contact with the measuring points and read the
measured value on the digital display.