Device description
4.3 Measuring ranges
5191 / 07/2021 en BENNING CM 2-1 25
4.3.5 Diode test
Overload protection: 600V‑AC/ V‑DC
Measuring range Resolution Measuring accuracy
1.5V 0.001V ±(1.5%+ 5digits)
Table12: Diode test
• Open-circuit voltage: approx. 1.8V
4.3.6 Capacitance ranges
Requirements: Discharge the capacitors and apply the safety measuring lines according to the
marked polarity.
Overload protection: 600V‑AC/V‑DC
Measuring range Resolution Measuring accuracy
1μF 0.001μF ±(2.0%+ 5digits)
10μF 0.01μF
100μF 0.1μF
1mF 0.001mF
10mF 0.01mF
Table13: Capacitance ranges (μF)
4.3.7 Frequency ranges
Overload protection: 600V‑AC/ V‑DC, 400A‑AC/ A‑DC
Measuring range Resolution Measuring accuracy
50Hz 0.01Hz ±(0.5%+ 5digits)
500Hz 0.1Hz
5kHz 0.001kHz
50kHz 0.01Hz
Table14: Frequency ranges (Hz)
• Minimum sensitivity:
– V-AC range 1Hz… 10kHz: >6V
eff
– V-AC range 10… 50kHz: >20V
eff
– A‑AC range 1Hz… 10kHz: >6A
eff
• Minimum frequency: 1Hz