Instructions
34
11/ 2019
BENNING CM 10–PV
7.3.2 INRUSH function (inrush current measurement)
additional error for functions AAC and (exible AC current transformer)
± 3 % with regard to the specied measuring accuracy
Average time: 100 ms
Trigger current: > 0.5 Ae in the 60 A measuring range, > 5 Ae in the 600 A
measuring range
7.4 Resistancemeasuringranges(Ω),continuityanddiodetesting
Function Measuring
range
Resolution Measurement accuracy*
Resistance
600.0 Ω 0.1 Ω ± 0.9 % + 7 digits
6.000 kΩ 0.001 kΩ
± 0.9 % + 4 digits
60.00 kΩ 0.01 kΩ
600.0 kΩ 0.1 kΩ
Continuity 600.0 Ω 0.1 Ω ± 0.9 % + 7 digits
Diode 1.500 V 0.001 V ± 0.9 % + 4 digits
* Before the measurement, a possible oset must be determined by short-cir-
cuiting the measuring leads and subtracting them from the measured value.
Overload protection: 1000 VAC/DC
Max. open circuit-voltage: approx. 1.8 V
The built-in buzzer sounds when resistance is less than 20 Ω to 200 Ω.
Response time: < 100 ms
Acoustic indication: 2.7 kHz
7.5 Capacity ranges (µF)
Conditions: Capacitors discharged and connected in accordance with the pola-
rity stated.
Measuring range Resolution Measurement accuracy
100.0 µF 0.1 µF
± 1.9 % + 4 digits
1000 µF 1 µF
Overload protection: 1000 VAC/DC
7.6 Frequency ranges (Hz)
Measuring range Resolution Measurement accuracy
100.00 Hz 0.01 Hz
± 0.3 % + 5 digits1000.0 Hz 0.1 Hz
10.000 kHz 0.001 kHz
Overload protection: 1000 VAC/DC, 600 AAC/DC
Minimum sensitivity:
> 5 Ve for V AC range (1 Hz - 10 kHz)
> 8 Ae for A AC range (1 Hz - 1 kHz)
Min. frequency: 1 Hz
7.7 Temperature range (°C/ °F)
Overload protection: 1000 VAC/DC
Measuring range Resolution Measurement accuracy*
- 40 °C - +400 °C 0.1 °C ± (1 % + 22 digits)
- 40 °F - +752 °F 0.1 °F ± (1 % + 38 digits)
* The measuring accuracy of the K-type temperature sensor has to be added
to the specied measuring accuracy.
Wire temperature sensor (type K): Measuring range: - 60 °C to 200 °C
Measurement accuracy: ± 2 °C
The measuring accuracy applies to stable ambient temperatures < ± 1 °C. After