Data Sheet
Page 6
Document No: SJ-GU-TF02-02 Version: A00
5
Glufosinate
√
√
6
Azoles
√
√
7
Hymexazol
√
×
Surface
corrosion
8
Phenoxy
acetic acid
√
×
There is
corrosion, and
material
becomes soft
9
Trifluralin
√
×
There is
corrosion, and
material
becomes soft
Description: in the test, diluted 100 times refers to the ratio of pesticide and water of 1: 100, that
is, the proportion of pesticide is 1%. The proportion of pesticides in the test is higher than
common pesticide proportion (0.1%-1%). The test product is completely immersed in nine
pesticides for at least 48 hours. √ indicates that the material is completely free from corrosion,
and × indicates that the material is affected by corrosion.
6. Detection Angle
TF02’s signal acceptance angle is 3˚, so LiDAR’s detection angle is 3˚. The side lengths of
different distance detection ranges (the detection range takes on a square shape) are shown
as below in the table.
Detection Range (Acceptance Angle) Diagram
Distance
(m)
1
2
3
4
5
6
7
8
9
10
15
20
22
Detection
Range
Side
Length
(cm)
5
10
16
21
26
31
37
42
47
52
79
105
115
Note: the distance in the table stands for the vertical distance between LiDAR and the detected
object, with meter as unit. The side length unit of detection range is cm. Generally, the side
length of detected object should be greater than that of detection range. Only in this way the
LiDAR output data can be reliable. When the side length of detected object is less than that of
detection range, the LiDAR output data is not stable and the error increases.
Note: the side length of detecting range is not equal to the object resolution at the
corresponding distance.
7. Dimension and Product Specification