User's Manual
8
Chapter 5
Control Solution Test
Control solution is used to check the performance of the kit.
The meter kit should be check:
• When the meter and strip do not work properly.
• When the test result is unusual or inconsistent.
Control solution range is shown as follow:
Please check your strip vial label for exact range.
Perform a Control Test:
Insert an Unused Strip
˛
Press “M” Button (3 secs)
˛
Enter Control Solution Mode
(Shown CL on Screen)
˛
Shake the Control Solution Well
˛
Discard First Three
Drops
˛
Put One Drop onto a Clean Surface
˛
Touch the Control Solution with Strip
Sample Inlet
˛
Test Result Display after Count Down
˛
Compare the Result with the
Range Listed on Test Strip Vial
Note:
• Do not reuse the test strip.
• Marked open date on new open control solution.
• Do not use the meter if the control test is out of range.
• If the control test keeps result wrongly, please contact an authorized distributor.
• Control solutions are not included. Please contact an authorized distributor for purchasing.
98-164 mg/dL
5.4-9.1 mmol/L
EA13120901
Control Range LV1