(3) Enter the required indicator display
Type the required BR323 indicator Zero and
Full scale displays into the appropriate panels.
Clicking the Upload button will enter these
figures and they will appear on the graph.
(4) Entering other parameters
Clicking the Input / Output tab will enable the
Filter, Tag, Display Resolution, Damping Time
and Units of Measurement to be entered – see
6.4 steps 3, 4, 5, 6 & 9 and Fig 7.
(5) Downloading calibration to indicator
Ensure that the Connection panel shows that
the BR323 indicator is connected to the
personal computer then click the download
arrow on the toolbar.
After downloading is complete the left hand side
of the Configuration Programme Screen should
show the new calibration information and the
BR323 should have the required display. The
calibration information may be stored or printed
from the PC in the normal way.
6.7 Other non-linear curves
Other non-linear curves may be generated by
entering the number of breakpoints required and
manually keying the required x and y values into
the table. Alternatively, a comma delimited csv file
containing the custom curve data may be imported
using the procedure described in section 6.6.