VCO-CV Tuning Procedure Service Support Document
page 3 of 4
MS101 VCO & CV
Calibration Procedures
2.1 Internal CV Calibration
1) Press “KYBD + STEP” then power on device. All the sequencer LED will be lit up.
2) Press “Shift + STEP1” to start Internal CV calibration.
3) Waiting for the tuning to be nished.
4) All the sequencer LED will be lit up after nish.
2.2 Output CV Calibration
In this mode, the buttons has new function:
There are four test points A / B / C / D to be calibrated. Their target voltage is 0 V / 1 V / 3 V / 5 V respectively.
Button Name Function Button Name Function
HOLD/REST Select test point A STEP1 Select X1 (about ±1.5 mV)
RESET/ACCENT Select test point B STEP2 Select X2 (about ±7.5 mV)
ARP/SETEND Select test point C STEP3 Select X3 (about ±15 mV)
PATTERN/BANK Select test point D STEP4 Select X4 (about ±30 mV)
SHIFT — STEP5 —
PAGE — STEP6 —
PLAY/STOP — STEP7 —
REC Save STEP8 —
KYBD Decrease
STEP Increase