Calibration
Synthesizers and Samplers
DEEPMIND 6
page 7 of 9page 7 of 9
To calibrate the KEYBOARD AFTERTOUCH, please use the following procedure:
NOTE: During the calibration the GLOBAL, -/NO and +/YES switches will be ashing. If at any time you want to cancel the
calibration routine, you can press the -/NO switch. No previous calibration data will be changed if you do cancel the routine.
To conrm the calibration at each stage press +/YES as described below.
CAL AFTERTOUCH
In order to start the calibration routine, make sure that the '<' symbol on the CAL AFTERTOUCH line is highlighted
and press +/YES.
You will see the following display showing a bar representing the current AFTERTOUCH value (0% for no pressure).
Also on this screen you can see the stored calibration data above the bar and the current value below the bar:
Now apply pressure to any one of the keys (or multiple keys) until the value stops increasing.
You will see the following display showing the maximum position of the AFTERTOUCH and the associated value:
NOTE: To cancel the routine you can press the -/NO switch.
To conrm the CALIBRATION of the AFTERTOUCH press the +/YES switch: The calibration data will then be stored for
the AFTERTOUCH and you will be returned to the CALIBRATE menu. From here you can perform another calibration
from the menu, or press the GLOBAL switch again to return to the SYSTEM SETTINGS menu.