User Manual

Operating manual for the GT-400 appliance tester
BEHA-AMPROBE
During the compensation procedure, a test is conducted to verify
whether the test line is connected to the protective earth contact
of the test socket, and the resistance of the test line does not
exceed 2 Ohm.
The following error messages appear in the status line of the
menu:
„Attention! Connect test line to protective earth contact!“
or
„Attention! Resistance of test line > 2 Ohm!“
(Test line too long, or insecure or faulty plug connections)
Compensation is automatically re-started following rectification of
the errors. If this is not possible, the procedure can be aborted
by pressing the „START“ button.
A faulty compensation is then stored.
Please note: If a faulty compensation has occurred, the
measuring sequence cannot be started when the rotary switch
is in the „SKI“ position.
12