User Manual
Operating manual for the GT-400 appliance tester
BEHA-AMPROBE
R
PE
selection
Since the limit value for the protective conductor resistance
measurement depends on the length of the protective conductor
of the test object among other factors, the menu allows the limit
value to be set according to the length (see also Section 8.2.)
Since the measurement of the protective conductor R
PE
can be
performed in the Protection class I switch position and also in the
Cable test switch position, the menu allows definition of which
limit values are valid for each type of measurement.
10
After selecting the validity by briefly pressing the „START“ button
(< 2 s), the desired selection is confirmed by pressing the
„START“ button for a longer time (> 2 s), after which you then
reach the following menu:
Here, you select the length of the protective conductor of the
test object („START“ button < 2 s). The limit value resulting from
this selection is displayed in the upper status bar. The selected
limit value is then confirmed by pressing the „START“ button for a
longer time (> 2 s).
The maximum limit value that can be set is
≤≤
≤
≤
≤ 1 Ohm. The limit
value is not permanently stored. This means that the limit value is
reset to
≤≤
≤≤
≤ 0.3 Ohm when the device is manually switched off via
the rotary switch or when the device automatically switches off
after approx. 3 minutes.
Valid for protection
Valid for cable test
Valid for protection
class I and cable test