Instructions

Harmonics in mA, A (Absolute Value)
specified from 100 mA rms 50/60Hz (mA Range) or from 1A rms 50/60Hz (A Range),
max. Peak Voltage 850 mAss (mA Range) or 70 Ass (A Range)
Range Resolution Accuracy (50/60Hz)
1...10 Display in mA: ± 1 % rdg. ± 2 mA
Display in A: ± 1 % rdg. ± 0,3 A
11...20 0,01 mA / 0,1 mA / Display in mA: ± 7 % rdg. ± 2 mA
0,001 A / 0,01 A Display in A: ± 7 % rdg. ± 0,3 A
21...50 Display in mA: ± 15 % rdg. ± 3 mA
Display in A: ± 15 % rdg. ± 0,3 A
51...99 Display in mA: ± 35 % rdg. ± 3 mA
Display in A: ± 35 % rdg. ± 0,3 A
Harmonics in mA, A and V Measurement Range
(Total Harmonic Distortion THD %)
specified from 100 mA rms 50/60Hz (mA Range) or from 1 A rms 50/60Hz (A Range) or
from 80 V rms AC 50/60Hz (V Range), max. Peak voltage 850 mAss (mA Range) or 70 Ass
(A Range) or max. Peak Voltage 850 Vss
Range Resolution Accuracy
0,0...10,0 % 0,1 % ± 2 %
10,0...40 % 0,1 % ± 5 % rdg. ± 5 %
40...100 % 0,1 % ± 10 % rdg. ± 10 %
100...999,9 % 0,1 % ± 20 % rdg.
PEAK Value in mA, A and V Measurement Range
specified from 10 mA rms (mA Range) or from 0,5 A rms (A Range) or from 20 V rms AC
(V Range), max. Peak voltage 850 mAss (mA Range) or 70 Ass (A Range) or max. Peak
Voltage 850 Vss
Range Sampling time Accuracy
50 Hz 39 µs ± 5 % rdg. ± 30 D
60 Hz 33 µs ± 5 % rdg. ± 30 D