Service manual
Revision 2/18/05 3 - 27
Hardware Description
Data Sheet: Condor GLM65B Medical Switching Power Supply
Functional Description
The Condor GLM65B medical grade power supply is
used in the Endeavor CR and VikingQuest base units to
supply +5V, +15V and -15Vdc to the VIASYS boards
and modules.
Both the Line and Neutral AC input lines are fused for
medical use. A blown fuse is an indication of
catastrophic failure of circuit component(s). This is a
replacement item only.
Specifications
AC Input Voltage: 85-264Vac, 47-63Hz single phase
Output Power: Total continuous power output is 60W,
75W peak for 60s 10% duty cycle.
Output Voltages (unrestricted convection cooling)
1. +5Vdc, 7A max
2. +15Vdc, 2.5A max
3. -15Vdc, 2A max
Connector Pinouts
J1: AC In
J2: DC Out
Additional Information
http://www.condorpower.com/index.html
Condor GLM65B Switching Power Supply
Pin Signal Pin Signal
1 AC Line 2 AC Neutral
Pin Signal Pin Signal
1 +15Vdc 4 Common
2 +5Vdc 5 Common
3 +5Vdc 6 -15Vdc
J1
J2
Pin 1
Pin 1
Condor GLM65B Power Supply