Service manual
Neurodiagnostic Instruments Service Manual
6-18 Revised 2/18/05
12. If in an office environment, are there any appliances
in the room with the system? Yes: ___ No: ___
Toasters, microwaves, computers, lamps,
coffee machines, etc.
13. Unplug any non-critical care equipment in the O.R. Heat lamps, microscope, electric bed,
warmers, etc.
14. Move the system to a different room away from the
existing location.
Move the system to a different part of the
building.
15. If using the system in the operating room,
momentarily unplug the Bovie Knife.
Disconnect Bovie from the wall outlet.
Troubleshooting Noise: Environmental Checklist
Items to check Detail
Troubleshooting Noise: Electrode Application Checklist
Items to check Detail
1. If using a Concentric Needle Adapter, make sure it
has a good patient ground connection.
The ground wire on the adapter should have
a patient ground lead
2. Do you have a good patient ground connection to the
amplifier?
Use the largest ground pad electrode you
have. Do not use a needle electrode for
ground.
3. Verify that all electrodes and/or needles are inserted
tightly into the amplifier connectors.
Push in each connector at the amplifier
inputs.
4. Verify that all electrodes and/or needles are attached
correctly to the patient.
Reapply electrodes if necessary.
5. Bundle/braid all recording electrode leads together.
There should be no looping wires on the patient.
Helpful in reducing 50Hz/60Hz noise.
6. Make sure the stim transducer cables, stim electrode
leads and any power cords are separated from the
recording electrode leads.
If the recording electrode leads must cross
other leads/cables, make sure they cross
perpendicular to each other.
7. Use electrodes that are made of the same types of
metal and the same manufacturer.
Do not mix metals or manufacturers.
8. Swap the existing electrodes with new ones. Electrodes do wear out.
9. Try a different brand of electrodes to eliminate the
possibility of a common flaw in the electrodes.
Type, batch or style of electrode.
10. Remove the EEG grounding mat if one exists. May act as an antenna for noise.
11. Do not place the ground electrode on middle of
patient’s chest. Try the shoulder or thigh instead.
EKG residual waves look much like 60Hz.