Service manual
Troubleshooting
Revised 2/18/05 6-13
File Management
The Diagnostic software provides several menu buttons
for managing test reports.
New
The New button creates a new text file that
contains the results of the current test.
Open
The Open button opens a previously created
test results file.
Save
The Save button lets you save a newly
created test results file.
Print
The Print button lets you print a hardcopy
of the test results file that is currently open.
Diagnostic Error Codes
When the Diagnostic software detects a failure, it
provides an error code to pinpoint the failure mode. The
error code is presented in the form xxx-yyy, where xxx
represents the LRU and yyy is the result code for pass or
fail.
The following charts summarize the error codes:
Endeavor CR Error Messages
LRU Error
Code
Error Message
100 1 STIM Test Passed
100 2 STIM Test Failed
100 3 STIM Test Failed:
Base board is not connected
100 4 STIM Test Failed:
Software error
100 5 STIM Test Failed:
Software setting error
100 6 STIM Test Failed:
Auditory board is not connected
200 1 Amplifier Test Passed
200 2 Amplifier Test Failed
200 3 Amplifier Test Failed:
Amplifier is not connected
200 4 Amplifier Test Failed:
Amplifier settings error
200 5 Amplifier Test Failed:
Calibration error
300 1 Control Panel Test Passed
300 2 Control Panel Test Failed
(not connected)
400 1 Stim Pod Test Passed
400 2 Stim Pod Test Failed
(not connected or illegal pods
configuration)
500 1 Headbox Test Passed
500 2 Headbox Test Failed:
Incorrect headbox type
500 3 Headbox Test Failed:
Amplifier is not connected
500 4 Headbox Test Failed:
No headbox has been connected
700 1 Auditory Stimulator Test Passed
700 2 Auditory Stimulator Test Failed
(not connected)