Service manual

Troubleshooting
Revised 2/18/05 6-3
USB/FireWire Diagnostics
The Endeavor CR, VikingQuest and VikingSelect
XP-based systems feature diagnostic packages with a
common user interface. The diagnostics are started from
a desktop icon:
The diagnostic packages allow a non-technical system
user to identify and correct hardware malfunctions
associated with VIASYS-designed hardware.
Malfunctions may include loose or improper
connections, as well as faulty Least Replaceable Units
(LRUs).
The diagnostics run in two modes.
The AutoTest mode runs without user interaction,
and provides a short Pass/Fail report for each tested
LRU.
The Manual test mode requires user interaction.
Individual tests are provided for the headbox,
control panel, auditory stimulator and electrical
stimulator.
To start the Diagnostics, click on the Diagnostics icon,
or select Start>Programs>Nicolet>Tools>Diagnostics.
A Caution window appears, reminding you to
disconnect all patient electrodes to the patient before
proceeding. Click on OK to close the window.
The Diagnostics window opens, displaying a menu bar,
an instruction panel, and a results panel. Each test
provides on-screen instructions.
The bottom of the page shows the diagnostic menu bars
for the Endeavor CR, VikingQuest, and VikingSelect
respectively.
Endeavor CR VikingQuest VikingSelect
Instruction Panel
Results Panel
Diagnostics Window
Diagnostics Menu Bars