Service manual
Neurodiagnostic Instruments Service Manual
5 - 10 Revision 2/18/05
Auditory Stim Verification
Visual Stim Verification
HTML Report Generation
Microsoft Word Report Generation
7. Verify that there are signals moving on each
channel in the Electrical Panel, and
averaging is taking place.
8. Place the bar electrode on your wrist or over
the thumb muscle. Carefully adjust the
Stimulus Intensity 1 knob until you feel the
stimulus, then readjust the output to zero.
9. Move the bar electrode to position 4 of the
SP-1 Stimulator Pod. If necessary, click on
the Start Average button to continue
averaging. Carefully adjust the Stimulus
Intensity 2 knob until you feel the stimulus,
then readjust the output to zero.
10. Move the bar electrode to position 7 of the
SP-2 Stimulator Pod. If necessary, click on
the Start Average button to continue
averaging. Carefully adjust the Stimulus
Intensity 2 knob until you feel the stimulus,
then readjust the output to zero.
11. Move the bar electrode to position 10 of the
SP-2 Stimulator Pod. If necessary, click on
the Start Average button to continue
averaging. Carefully adjust the Stimulus
Intensity 2 knob until you feel the stimulus,
then readjust the output to zero.
12. Click on the Stop Preview button and
disconnect the electrodes.
Step Action
1. Verify that the headphones are plugged in.
2. Click inside the Auditory panel to select it.
3. Click on the Auditory button to display the
Auditory window. Set the Left and Right
intensities to 65dB.
4. Click on the Start Preview button.
5. Click on Start Average button.
6. Verify you can hear a click stimulus in both
ears.
7. Click on the Stop Preview button.
Step Action
Step Action
1. Click inside the Visual panel to select it.
2. Click on the Start Preview button.
3. Click on the Start Average button.
4. Verify the goggles flash in accordance with
the screen display choices.
5. Click on Stop Preview.
Step Action
1. From the menu bar, select File > Report >
Generation to Html.
2. Highlight the
Endeavor_Default_Portrait.set file.
3. Click on Open.
4. Verify that the selected Endeavor report
opens in Microsoft Internet Explorer.
5. Exit Internet Explorer.
Step Action
1. Click on File > Report > Generation to
Word.
2. Highlight the
Endeavor_Default_Portrait.set file.
3. Click on Open.
4. Verify that the selected Endeavor report
opens in Microsoft Word.