Service manual

Revision 2/18/05 5 - 3
System Procedures
System Verification Checklist
Use this checklist and the following functional procedure to verify the system is working properly. Most steps require
no access to the interior of the instrument, and may be performed by the system operator. Internal desktop computer
checks must be performed by a qualified and trained Biomedical Engineer.
System Inspection & Cleaning
Check board seating & internal cables
(desktop systems only)
Clean/vacuum dust from ventilation slots
Check external cables
Clean instrument exterior/accessories
Check fans
Computer
System boots properly
Check Device Manager
Check event logs
Run Check Disk
Run Defrag
USB/FireWire Diagnostics
AutoTest
Board Report
Amplifier/headbox test
Auditory stimulator test
Electrical stimulator test
Control panel test
Miscellaneous Functions
LED goggles
Foot switch
Reflex hammer
Trigger In
Trigger Out
EMG speaker
Temperature read
Impedance Limited reading (VS only)
Application Verification Tests
(Perform one or more of the following)
VikingQuest and VikingSelect
Auditory Evoked Potential
Visual Evoked Potential
Somatosensory Evoked Potential
Motor Nerve Conduction
Sensory Nerve Conduction
SPA EMG (surface electrode only)
Report Generation (HTML or MS Word)
Report printing
Endeavor CR
Endeavor Application Checks
System Owner:
System Serial Number:
Software Version: