Specifications

ESU A
ESU A
N A LY Z E R S
N A LY Z E R S
& A
& A
C C E S S O R I E S
C C E S S O R I E S
H
H
1-888-BC-FOR-ME
(888-223-6763)
A. MicroMed MM-5kV Portable Electrosurgical Instrument
Insulation Detector . . . . . . . . . . . . . . . . . . . . . . $1550
McGan Technology’s MicroMed MM-5kV electrosurgical insula-
tion defect detector is a compact, hand held, battery operated
device. The MM-5kV may prevent inadvertent burns and exter-
nal skin injury which can occur during electrosurgical proce-
dures. Electrosurgical instruments may have defects (pin-
holes) in the insulation - electrosurgical instrument insulation
failure. Insulation failure happens over time, through normal
surgical use. The insulation failure provides an alternate path-
way for the current to leave the instrument's electrode and can
result in a burn. Using an electrosurgical test equipment main-
tenance program can greatly reduce the number of inadvertent
burns during electrosurgical procedures.
Application: in Facilities that do not use a large number of ESIs
over a short period of time
Light weight portable unit
Reusable / Interchangeable Brush or Ring electrode
Simple operation, easy to read LCD display
Maintains applied test voltage with constant current source
Full test current at low voltages
Limited Output current for operator safety
LED displays alarm and battery charge
Accessories:
Ground (Green) wire with Alligator clip
Ring and Brush Electrode
B. MicroMed PD-8K Portable Electrosurgical Instrument
Insulation Detector . . . . . . . . . . . . . . . . . . . . . . $2686
McGan McGan Technology's MicroMed PD-8K electrosurgical
insulation defect detector is a portable, battery operated
device. It is used to detect and locate defects such as pinholes
and cracks in protective coatings applied over conductive
instrument's (ESI) surfaces.
Application: in Facilities that use a large number of ESIs over a
short period of time
Operational during battery re-charging
Simple operation, easy to read LCD display
Multiple alarms to indicate failure
Maintains applied test voltage with constant current source
Limited output current for operator safety
Regulated DC output, adjustable from 0 to 8 kV
LCD displays provides voltage and battery charge indication
Accessories:
Ground (Green) wire with Alligator clip and Green port
Ring and brush electrode
HV (Red) wire
Optional Accessories for A or B:
C. MMTRI-0022A Tri-Hole Electrode . . . . . . . . . . $110
MMWTU-0031 Wire continuity testing unit . . . . . . $68
D. MMRTF-0045 RTF Test Station . . . . . . . . . . . . $310
Optional Accessories for A:
MMRWP-0006 HV Red wire with mini-handle . . $115
5VEBP-0016 External rechargeable battery pack . . . . $525
5VACA-0015 110 volt AC Charger . . . . . . . . . . . $88
A
B
C
D