User manual
35
RESPIRATION SIMULATION (NIBP-1020 & NIBP-1030)
RATE ACCURACY
± 1%
IMPEDANCE
DELTA
3.0 Ω ± 10%
BASELINE
1000 Ω ± 5%
TEMPERATURE SIMULATION (NIBP-1030 ONLY)
SENSOR TYPE
YSI Series 400 and 700
POINTS
0, 24, 30, 35, 37, 40, 42 °C
(32, 75.2, 86, 95, 98.6, 104, 107.6 °F)
ACCURACY
± 0.1 °C
ELECTRICAL
AC POWER ADAPTER
12 VDC, 500 mA
BC20-40613 (USA)
BC20-40614 (Euro)
BATTERY (OPTIONAL)
Rechargeable NiMH pack
(Not user serviceable)
BATTERY RUNTIME
500 Test Cycles between charges
PHYSICAL & ENVIRONMENTAL
DISPLAY
128 X 64 Pixels Graphical LCD,
White LED Backlight
SIZE
NIBP-1000 through
NIBP-1030
7.70 x 5.36 x 4.07 Inches
(195.6 x 136.1 x 103.4 mm)
NIBP-1030-BE
7.70 x 6.15 x 4.07 Inches
(195.6 x 156.2 x 103.4 mm)
WEIGHT
< 3.5 Lbs (1.6 kg)
CONSTRUCTION
ENCLOSURE
ABS Plastic
OVERLAY
Back-Printed Lexan
OPERATING RANGE
15 to 40 °C
(59 to 104 °F)
STORAGE RANGE
-20 to 65 °C
(-4 to 149 °F)