User manual
27
LEAK TEST: (NIBP-1030 Only)
Use the key to cycle through the Test Modes until Leak Test appears. The
purpose of this function is to provide a standard pressure leak test. The pressure input is
connected to the system that is to be monitored. Pressure is then applied to the system.
The Leak Test is initiated by pressing the key. The unit traces and displays the
pressure drop and the time since the test was initiated. The unit also calculates and
displays the leak rate in mmHg per minute.
NOTE: While in this mode, the NIBP simulation output is disabled.
The displays will resemble the following examples: (Typical Screens)
(Pressure Only Screen)
(Output Waveform Screen) (Pressure Graph Screen)
LEAK TEST
PRESSURE
PEAK PRESSURE
Drop in mmHg
Time in min:sec
Leak Rate in
mmHg/minute