User manual
26
ARRHYTHMIA SEQUENCE: (NIBP-1030 Only)
Use the key to cycle through the Test Modes until Arrhythmia Sequence appears.
The purpose of this function is to provide a timed sequence of some of the more common
Arrhythmias. The test will continually cycle through the 6 arrhythmias in the sequence listed
below. A count down timer is displayed to indicate the time still remaining in each step.
NOTE: While in this mode, the NIBP simulation output is disabled.
The displays will resemble the following examples: (Typical Screens)
(Pressure Only Screen)
(Output Waveform Screen) (Pressure Graph Screen)
Step
Display
Total Time
ECG Output
1
Single PVCs
18 sec
11 NSR, 1 PVC, 11 NSR, 1 PVC
2
Pair of PVCs
9 sec
10 NSR, Pair of PVCs
3
Single PVCs
18 sec
11 NSR, 1 PVC, 11 NSR, 1 PVC
4
Run of 5 PVCs
12 sec
11 NSR, Run of 5 PVCs
5
Vtach
30 sec
15 NSR, VTach
6
Vfib
30 sec
15 NSR, VFib
7
Asystole
30 sec
15 NSR, Asystole
Single PVCs
Pair of PVCs
Single PVCs
Run of 5 PVCs
Vtach
Vfib
Asystole
ARRHYTHMIA
PRESSURE ENVELOPE
OFFSET
This line is only displayed
if a non-zero value is
programmed into the
Systolic &/or Diastolic
Shift. (See Setup for
more information).
PRESSURE
(NIBP Simulation disabled,
this is pressure of the static
volume only)
PEAK PRESSURE
(NIBP Simulation disabled,
this is peak pressure of the
static volume only)
TIMER