User manual

25
ALARM TEST:
Use the key to cycle through the Test Modes until Alarm Test appears. The
purpose of this function is to help test the alarms of the monitor under test. The ECG
output will alternate from a 47 BPM NSR with apnea respiration (0 brpm) to 160 BPM NSR
with 80 brpm respiration. The time interval between alternations is 30 seconds. A count
down timer is displayed to indicate the time still remaining for the current output.
NOTE: While in this mode, the NIBP simulation output is disabled. Pressure readings in
the screens below are only manometer-type readings of static pressure applied.
The displays will resemble the following examples: (Typical Screens)
(Pressure Only Screen)
(Output Waveform Screen) (Pressure Graph Screen)
PEAK PRESSURE
(NIBP Simulation disabled,
this is peak pressure of the
static volume only)
ALARM TEST
PRESSURE
(NIBP Simulation disabled,
this is pressure of the static
volume only)
PRESSURE
ENVELOPE OFFSET
This line is only
displayed if a non-zero
value is programmed
into the Systolic &/or
Diastolic Shift. (See
Setup for more
information).
Heart Rate 47/160
Respiration Rate 0/80
TIMER