User manual

22
There are 5 selectable NIBP Basic Test Modes; Adult, Adult w/pace, Adult High, Adult Low
and Neonatal. The displays will resemble the following examples:
NOTE: The screens for the first test will have the individual components labeled. The
component labels for subsequent tests are the same.
Adult 120/80
(Typical Screens)
(Pressure Only Screen)
(Output Waveform Screen)
(Pressure Graph Screen)
PEAK PRESSURE
(Optional)
PRESSURE
PRESSURE ENVELOPE
OFFSET
This line is only displayed
if a non-zero value is
programmed into the
Systolic &/or Diastolic
Shift. (See Setup for
more information).
PEAK PRESSURE
(Optional)
NIBP TEST
PEAK PRESSURE
(Optional)
PRESSURE
Type Adult 120/80
Heart Rate 80
Respiration Rate 35
(Optional)
NIBP TEST
PRESSURE
OUTPUT
WAVEFORM
(as generated by the
NIBP-1000 Series)
Dependant on
pressure
PRESSURE
GRAPH
2mmHg per
pixel
Type Adult 120/80
Heart Rate 80
Respiration Rate 35
(Optional)
Type Adult 120/80
Heart Rate 80
Respiration Rate 35
(Optional)
NIBP TEST