User manual

21
Depending on the model, there are up to 9 Basic Test Modes and 3 Sub-Test Modes. This
section will walk through each of the tests and their basic operation.
The Main Tests are accessible with a single key. The key will scroll through the
following tests in a continuous loop:
Adult 120/80
Adult 120/80 w/Pace
Adult High 190/120
Adult Low 80/40
Neonatal 70/40
Alarm Test
Arrhythmia Seq
Leak Test
Manometer
BASIC TEST MODES
NIBP:
The first 5 test modes deal with various NIBP setups. To run an NIBP simulation, the cuff
and monitor are connected to the pressure input. Then the measurement is initiated by the
monitor and the NIBP Simulator will output the proper waveform based on the cuff
pressure provided by the monitor and the selected simulation.
The NIBP output mode can be changed by pressing the key. Once the desired
operating mode is selected, the output will automatically begin when the correct pressure is
detected.
RUNNING A TEST